Johnson & Johnson Vision Care QRB-08-2017 - 1-DAY ACUVUE® MOIST® Brand Contact Lenses
We received a limited number of reports of “stinging of the eyes” when the lens is inserted. Analysis of the solution confirmed ...
-
Medline FSN17 06 - MED-SOFT 3-liter soft liners
Aan het einde van de afzuigprocedure kan bij het loskoppelen van de rode vacuümslang, onmiddellijk of later, de dop losschieten.
-
Autochair Ltd AC-FSN-001-MHRA - SmartBase
The combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions in the intended ...
-
Roche Diagnostics FSN SBN-RMD-2017-009 - cobas® MRSA/SA Nucleic acid test
Roche Diagnostics is advising that you discontinue the use of and discard any remaining inventory of the cobas® MRSA/SA Nucleic ...
-
Maquet FSCA-2017-10-16 - Heater-Cooler Units HCU 20 and HCU 30
Maquet Cardiopulmonary as well as different National Competent Authorities have received isolated reports indicating bacterial ...
-
Siemens CT053-17-S - SOMATOM Force, SOMATOM Definition AS, DS, Edge, Flash
This letter is to inform you about a potential risk of unnecessary radiation exposure due to a software issue we found in the ...
-
Sanquin Reagents FSN 103549 - Screening panel
Bij kwaliteitsonderzoek is gebleken dat cel van deze panels (donorcode 384549) op het bijbehorende antigram ten onrechte staat ...
-
Johnson & Johnson Vision Care QRB-09-2017 - ACUVUE® OASYS® for ASTIGMATISM
We received a limited number of confirmed reports of a lens being ‘off power’ which the consumer may recognize as not accurately ...
-
Ion Beam Appplications (IBA) PR-83365 - ProteusPLUS and ProteusONE
It has been observed during testing that there could be clinically non-negligible amount of dark current in Pencil Beam Scanning ...
-
Teleflex EIF-000216 - Percuvance® percutaan operatiesysteem
Teleflex roept deze producten terug omdat de bek van de toolpunt kan breken, hetgeen ertoe leidt dat een fragment van de toolpunt ...