FSN-BC-020988, Biocartis NV, Idylla™ Instrument Test
In the context of a complaint investigation, Biocartis has identified a set of Instruments manufactured between September 2021 ...
-
Elekta, Inc., FCA-IMS-0044, Monaco RTP System
Een reeks stappen die door de gebruiker worden uitgevoerd, bestaande uit een initiële optimalisatie, uitbreiding van een ...
-
Eakin Surgical Limited, ESL-2301, Diathermy Suction
Testing of the Diathermy Suction Devices has revealed non-compliance with IEC 60601-2-2-2017 standard: Medical electrical ...
-
Philips Medical Systems DMC GmbH, 2021-PD-DXR-007/2022-PD-DXR-015, CombiDiagnost R90 ProxiDiagnost N90
Philips has previously notified you of issues relating to potential component damage due to short circuit (CombiDiagnost R90, ...
-
Datascope Corp, 2249723-01/24/2023-003-C, Cardiosave Intra-Aortic Balloon Pumps
There have been reported failures of the high pressure helium regulator which may cause a helium leak in the Cardiosave Hospital ...
-
CareDx AB, A-2023-007, Olerup QTYPE 11
De mix op positie K21 in het O560-kanaal op de QTYPE-plaat lot E061 is ontworpen om het DPB1*17:01-allel te amplificeren, maar is ...
-
FSN-FSCA29, HYPHEN BioMed, BIOPHEN TM Protein C LRT
The R2 of the products listed above contains Cesium Chlodride. Following information from our supplier, this substance has ...
-
Thermo Fisher Scientific, FSN-2023-001, Thermo Scientific Oxoid MacConkey Agar wout Salt
Continued use of this lot may result in delay to patient treatment
-
Datascope Corp., 2249723-01/24/2023-002-C, Cardiosave Intra-Aortic Balloon Pumps
As part of an ongoing investigation into unexpected shutdown of the Cardiosave IABP, the Datascope/Getinge Investigation team ...
-
Rapport Lokale teams, Toegang in Middelburg
De Inspectie Gezondheidszorg en Jeugd (hierna: inspectie) voerde in oktober 2022 aangekondigd toezicht uit bij Toegang In ...