Recall of Procedure Packs- kits-: IMP Tumorstent, Innoflex Mono-J, Innoflex Ureteral stent, Innosoft Mono-J, Innosoft Ureteralstent, Innostar Ureteralstent, Percudrain 2 step nephrostomy, Percudrain 3, Percusoft 3 step drainage, Percusoft 3 step nephrostomy set, Balloon 2 step Nephrostomy

Certain Procedure Packs assembled by BS Medical Tech Industry (BS-MTI) contain guidewires manufactured by EPflex Feinwerktechnik GmbH. The sterile packaging of these guidewires failed the microbial barrier test after five years of real-time aging. Consequently, the ability of the guidewire packaging to maintain sterility throughout its claimed five-year shelf life can no longer be assured. The issue concerns the sterile packaging of the EPflex guidewires, not the sterile packaging of the other components included in the BS-MTI Procedure Packs

Fabrikant: BS MEDICAL TECH INDUSTRY
Product: Procedure Packs
datum FSN: 27-10-2026
 

Waarschuwing van fabrikant

Dit bericht is een waarschuwing van de fabrikant. Na het op de markt brengen van een medisch hulpmiddel is de fabrikant verplicht het hulpmiddel te volgen en waar nodig te verbeteren. Wanneer sprake is van een risico voor de gebruiker moet de fabrikant actie ondernemen. De fabrikant informeert gebruikers en de inspectie over deze actie met een waarschuwing (Field Safety Notice). Zie ook waarschuwingen medische hulpmiddelen. Dit hoort bij het toezicht op medische technologie.