Olympus received complaints for the ViziShot 2 FLEX (19G) devices in which the hypotube component ejected from the device or plastic components detached during use. Investigation has identified that device heat-shrink material degradation and use errors can cause hypotube component ejection. Additionally, investigation has identified that the device heat-shrink material (which seals the needle) degrades during clinical use and may result in difficulty in extracting or expelling samples, fluid leakage, impaired needle deployment or retraction, or breakage of device components. To mitigate these risks, Olympus is removing all ViziShot 2 FLEX (19G) devices from the market.
Fabrikant: Gyrus ACMI, Inc. (Olympus Surgical Technologies America)
Europees vertegenwoordiger: Olympus Europa SE & Co. KG
Product: ViziShot 2 FLEX (19G)
Datum FSN: 12 januari 2026
Waarschuwing van fabrikant
Dit bericht is een waarschuwing van de fabrikant. Na het op de markt brengen van een medisch hulpmiddel is de fabrikant verplicht het hulpmiddel te volgen en waar nodig te verbeteren. Wanneer sprake is van een risico voor de gebruiker moet de fabrikant actie ondernemen. De fabrikant informeert gebruikers en de inspectie over deze actie met een waarschuwing (Field Safety Notice). Zie ook waarschuwingen medische hulpmiddelen. Dit hoort bij het toezicht op medische technologie.