Documenten - Waarschuwingen van fabrikanten medische hulpmiddelen
30 publicaties over Waarschuwingen van fabrikanten medische hulpmiddelen
- Datum
- Relevantie
-
Mölnlycke Health Care AB, Mölnlycke Trays
The reason for this FSCA is that Fairmont has identified that the white suction button of the device doesn’t return to the rest ...
-
Channel Medsystems, Inc., 3012018285.04242024.C.00001, Cerene Cryotherapy Device
Error code 003 was not listed in certain rows of the Cerene Cryotherapy Device Instructions for Use “Troubleshooting” section ...
-
Intuitive Surgical, Inc., ISIFA2024-02-R, SureForm45/60, Cannula Seal, Arm Drape, Obturator
Intuitive is tot de ontdekking gekomen dat er bepaalde partijen verzonden zijn die over hun officiële vervaldatum waren.
-
Unimax Medical Systems, Inc., FCA-EP24001, Detachable EndoRetrieval Pouch
The mechanism of the listed article number operates in a way that the tube within detaches during the removal process. If the ...
-
Olympus Medical Systems Corporation, QIL FY24-EMEA-40-FY24-OMSC-36, Single Use Electrosurgical Knife
Olympus's investigation has identified that deterioration of the cutting knife can contribute to tip breakage during use. ...
-
Intuitive Surgical, Inc., ISIFA2024-02-R, SureForm45/60, Cannula Seal, Arm Drape, Obturator
Tot onze spijt moeten wij u mededelen dat Intuitive zich ervan bewust is geworden dat bepaalde partijen van de volgende producten ...
-
Covidien IIc, FA1398, Auto Suture™ Structural Balloon Trocar, Auto Suture™ Blunt Tip Trocar
We hebben meldingen ontvangen van voorvallen met betrekking tot de Covidien Auto Suture™ structurele trocar met ballon en Auto ...
-
CooperSurgical, Origio® Sperm Wash Medium
CooperSurgical has internally become aware of a non-conformity of Origo® Sperm Wash Media related to the low concentration of ...
-
Gyrus ACMI, Inc., SG PK CUTTING FORCEPS, 5MM, 33CM. Olympus
The jaw may fracture prior to the procedure during the inspection instructed per the IFU, or during the procedure.
-
KARL STORZ SE & Co. KG-23-0015, BEYER Antrum Punch, Antrum Punch, Sphenoid Punch, Sphenoid Punch, Sphenoid Punch, Galea Spring Hook, Uvula Retractor, Optical Biopsy and Grasping Forceps, Grasping Forceps, Optical Scissor, Working Insert with steering lever
It was found that there is not insufficient evidence to show that the reprocessing method of the products was adequately ...