Cepheid, 24FAR003, Xpert Carba-R
Vals-positieve patiënt testresultaten kunnen leiden tot onnodige contactvoorzorgsmaatregelen om verspreiding van de infectie te ...
-
Stago, RC-24-0001, Star Max. Sta-R Evolution. Compact Max
Stago has investigated and confirmed a potential risk of pipetting close to or into the packed red blood cells when using these ...
-
Randox Laboratories Ltd, REC731, CALIBRATION SERUM LEVEL 3 (CAL 3)
Randox Laboratories is erachter gekomen dat het anorganisch fosfaat in kalibratieserum niveau 3, CAL2351, een negatieve afwijking ...
-
Siemens Healthcare Diagnostics Inc., ACHC 24-03, Atellica CH Iron3
Siemens Healthineers has confirmed the potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica® CH and ...
-
Becton, Dickinson and Company, IDS-23-4910-B, BD BACTEC™ MicroMGIT Calibration Vial
In november 2023 startte BD een herstelactie voor de BD BACTEC™ MicroMGIT™ - calibration vials met betrekking tot lage ...
-
Randox Laboratories Ltd., REC718, CYSTATIN C CALIBRATOR (CYSC CAL)
Randox Laboratories is erachter gekomen dat de Cystatin C Calibrator uit de CYS 2699-serie een negatieve afwijking vertoont in ...
-
Life Technologies Holdings Pte Ltd., PR# 756842, on PGM™ Dx Instrument System
This communication is to inform you that Thermo Fisher Scientific has recently completed penetration testing on the Torrent ...
-
Siemens Healthcare Diagnostics Inc, POC 24-007, Atellica UAS 800 Analyzer
Siemens Healthcare Diagnostics Inc. has confirmed an issue that has the potential to misidentify samples when ALL the following ...
-
Abbott Molecular, Inc., FA-AM-MAR2024-298, Alinity m System
The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present ...
-
Hologic Inc., FA-00242, Panther Fusion SARS-CoV-2/Flu A/B/RSV Assay
While testing the Panther Fusion SARS-CoV-2/Flu A/B/RSV assay on Panther Fusion System, there is a potential risk of a false Flu ...